Gemini Users Manual 90
If this happens immediately after start-up, limits may be incorrectly set. Cold start
the mount, and check the limits. Slew to the places where the limits should be and
select “SetupMount ParametersSet Safety Limit” for both the east and the
west limits.
7.1.12 Autoguider Calibration Problems
Be certain that you are in Photo Mode or All Speed Mode. The autoguider port
will only work in these two modes. Make sure you have not left Gemini in Menu
Mode, and that you are using a tracking rate that is appropriate for your
observation.
Make sure you have the correct cable to connect the autoguider to the Gemini. Be
sure to use an Optocoupler unit or Relay Box (with appropriate cable) if you have
an autoguider without relays and an older Gemini (motor cables use RJ-45
connector). If your Gemini unit uses DIN connectors for the motors, you can
connect TTL output autoguiders directly to Gemini. Refer to Appendix 8.6 for
more details.
If RA calibration works but Dec. does not, check the backlash in the Dec. gear.
Set the TVC or increase the autoguider’s hysteresis setting.
7.1.13 Gemini Displays “Autoguider Error”
This message indicates that Gemini received contradictory signals (RA+ and RA-
or DEC+ and DEC-) at the same time. If this happens occasionally, it can be
ignored. Otherwise check the cable from the autoguider to the Gemini and keep it
as short as possible. Do not use a twisted cable; it could induce spurious signals
from one wire to the other.
7.1.14 Serial Communications Problems
Ensure that you are using the correct cable between Gemini and the PC or GPS
receiver. Refer to Appendix 8.5 for details.
Start Gemini in Debug Mode and look at the information buffer to see if data is
actually being received from the PC. Refer to section 7.2 below.
If you are using a PC with a fast CPU and buffered serial I/O circuitry, the
microcontroller of the Gemini system may not be fast enough to receive and store
all characters sent to it. Check the setting of the communication port of your PC
and disable the 16550A serial controller’s FIFO. You can try to use a smaller
FIFO send buffer first.
See also section 6.1.3.
7.1.15 "CMOS reset" Message upon Startup or Incorrect Date/Time
The internal Li battery was exchanged or is low. Power off the Gemini and open
the Gemini electronics unit. Replace the internal battery with a fresh one. Older
Gemini units use a CR2032 Li battery, while newer units use a CR2354 Li
battery. Be sure to replace the battery with the same type that came with your
Gemini. After you power up the Gemini again, you will see this message. Use